Notification Specification
On an nPOC-BB (Gamma) unit running firmware v3.5, the LED notification lights will notify the user according to the specification in the table below.
| Status LED | Status | Battery LED | Battery | Description | User action |
|---|---|---|---|---|---|
| Off | Off | Module is off for storage or transport | Turn on the power switch before starting a run | ||
| Off | Any | Module is ready | Can load STC in the module | ||
| Green(flashing) | Any | Run is in progress | Wait until the run is completed | ||
| Green | Any | Run is completed | Remove Sample Tube for next user step | ||
| Yellow(flashing, 3 s) | Any | ST/TC is not loading correctly | Remove and reload ST/TC | ||
| Yellow | Any | Device is not ready | Do not load sample tube/test consumable | ||
| Red(flashing, 3 s) | Any | Error encountered. LED will transition to solid red | No action | ||
| Red | Any | Run is stopped with error. Run is invalid | Remove Sample Tube from the unit and do not proceed to next user step | ||
| Yellow/Purple (alternating) | Any | Critical device error | Request for service from a qualified technician | ||
| Any | White | Battery is >90% | No action | ||
| Any | Yellow | Battery is sufficient for 1 run | Recharge battery as soon as possible | ||
| Any | Red | Battery is insufficient; Module will not allow a run | Recharge battery | ||
| Any | Blue(breathing) | Module is charging | No action | ||
| Blue(breathing) | Blue(breathing) | Module is in mass storage mode | Access log file or changing config parameters via PC |