Risk Evaluation

Overview

The nPOC-BB was originally developed at GHL as the front-end of a prototype diagnostic test system called NAATOS. Therefore, the design inputs for the nPOC-BB are a subset of the design inputs for NAATOS, and the Risk Evaluation for the nPOC-BB is excerpted from the larger NAATOS Risk Evaluation.

You can also download a tabular version of this Risk Evaluation Table in the *.csv format.

Lookup Matrices

Probability

P1 (Probability of the Hazardous Situation Occurring) × P2 (Probability of the Hazardous Situation Leading to Harm) = Probability

P1 \ P2 1 - Improbable 2 - Remote 3 - Occasional 4 - Probable 5 - Frequent
1 - Improbable 1 - Improbable 1 - Improbable 2 - Remote 2 - Remote 3 - Occasional
2 - Remote 1 - Improbable 2 - Remote 2 - Remote 3 - Occasional 3 - Occasional
3 - Occasional 2 - Remote 2 - Remote 3 - Occasional 4 - Probable 4 - Probable
4 - Probable 2 - Remote 3 - Occasional 4 - Probable 4 - Probable 5 - Frequent
5 - Frequent 3 - Occasional 3 - Occasional 4 - Probable 5 - Frequent 5 - Frequent

Risk Level

S (Severity) × P (Probability) = Risk Level

P \ S 1 - Negligible 2 - Minor 3 - Serious 4 - Critical 5 - Catastrophic
1 - Improbable Low Low Low Low Medium
2 - Remote Low Low Medium Medium Medium
3 - Occasional Low Medium Medium Medium High
4 - Probable Low Medium Medium High High
5 - Frequent Medium Medium High High High

Risk Evaluation Table

Name Sequence of Events Hazardous Situation Harm Severity P1 (Probability of the Hazardous Situation Occurring) P2 (Probability of the Hazardous Situation Leading to Harm) Ptotal = P1 x P2 Initial Risk Level Initial Risk Acceptability
SCC contaminates environment with sample
  1. SCC collects sample from an infectious patient
  2. SCC touches objects in environment before transfer into STC or handle touches objects in environment after separation from head in STC
Person exposed to sample Disease transmission 3 - Serious 2 - Remote 3 - Occasional 2 - Remote Medium Unacceptable
SCC is reused
  1. SCC collects sample from an infectious patient
  2. SCC collects sample from a different patient
Person exposed to sample Disease transmission 3 - Serious 1 - Improbable 5 - Frequent 3 - Occasional Medium Unacceptable
STC contaminates environment with sample
  1. SCC collects sample from an infectious patient
  2. SCC transfers sample into STC
  3. STC spills sample onto workspace or person
Person exposed to sample Disease transmission 3 - Serious 2 - Remote 3 - Occasional 2 - Remote Medium Unacceptable
STC is reused
  1. STC with remnants of sample from an infectious patient is re-opened
Person exposed to sample Disease transmission 3 - Serious 1 - Improbable 1 - Improbable 1 - Improbable Low Acceptable
SPM contaminates environment with sample
  1. a. SPM containment fails during operation or b. STC containment fails inside SPM during operation
  2. SPM shakes infectious sample into the environment
Person exposed to sample Disease transmission 3 - Serious 2 - Remote 4 - Probable 3 - Occasional Medium Unacceptable
SPM does not inactivate sample
  1. a. SPM is mis-calibrated or b. SPM software fails or 1c. SPM electronics fail
  2. SPM appears to heat and shake but fails to inactivate sample in STC
  3. STC allows transfer of infectious sample to person
Person exposed to sample Disease transmission 3 - Serious 1 - Improbable 2 - Remote 1 - Improbable Low Acceptable
SCC is contaminated with manufacturing chemicals
  1. SCC is contaminated with chemicals during manufacture
  2. SCC is used to collect sample from person and transfers chemicals to sampling site on person
Patient exposed to chemicals Toxicity 3 - Serious 1 - Improbable 2 - Remote 1 - Improbable Low Acceptable
SCC is contaminated by environment
  1. SCC collects MTb from the environment
  2. SCC is used to collect a negative sample
Patient receives false positive result Incorrect diagnosis 3 - Serious 1 - Improbable 4 - Probable 2 - Remote Medium Unacceptable
SCC is never added to STC
  1. SCC is never added to STC
Patient receives false negative result Incorrect diagnosis 4 - Critical 1 - Improbable 4 - Probable 2 - Remote Medium Unacceptable
STC is contaminated by environment
  1. STC collects MTb from the environment
  2. STC is used to process a negative sample
Patient receives false positive result Incorrect diagnosis 3 - Serious 1 - Improbable 4 - Probable 2 - Remote Medium Unacceptable
STC is opened and left exposed to environment
  1. STC is left open until liquid reagents evaporate
Operator obtains invalid result Delayed diagnosis 1 - Negligible 1 - Improbable 3 - Occasional 2 - Remote Low Acceptable
STC is removed from SPM during operation
  1. STC with positive sample is removed early from SPM
  2. STC transfers insufficiently lysed sample to TC
  3. TC fails to amplify insufficiently lysed sample
Patient receives false negative result Incorrect diagnosis 4 - Critical 1 - Improbable 3 - Occasional 2 - Remote Medium Unacceptable
STC sample transfer into TC introduces air bubbles
  1. STC sample transfer introduces air bubbles into TC
  2. TC fails fluidically because of bubbles
Operator obtains invalid result Delayed diagnosis 1 - Negligible 2 - Remote 2 - Remote 2 - Remote Low Acceptable
STC sample transfer into TC is delayed
  1. STC sample transfer of a positive sample into TC is delayed
  2. STC sample degrades
  3. TC fails to amplify degraded positive sample
Patient receives false negative result Incorrect diagnosis 4 - Critical 1 - Improbable 2 - Remote 1 - Improbable Low Acceptable
STC sample transfer into TC leaks
  1. STC sample transfer into TC leaks
  2. TC fails fluidically because of insufficient fluid volume
Operator obtains invalid result Delayed diagnosis 1 - Negligible 1 - Improbable 3 - Occasional 2 - Remote Low Acceptable
STC leaks fluid into SPM
  1. STC leaks while in SPM
  2. SPM electronics are shorted by leaking fluid and catch fire
Operator exposed to fire Burn 3 - Serious 1 - Improbable 2 - Remote 1 - Improbable Low Acceptable
SPM batteries overcharge
  1. SPM batteries overcharge, heat up, and catch fire
Operator exposed to fire Burn 3 - Serious 1 - Improbable 4 - Probable 2 - Remote Medium Unacceptable
SPM batteries over-discharge
  1. SPM batteries over-discharge
  2. SPM can no longer hold a charge
Operator unable to use SPM Delayed diagnosis 1 - Negligible 1 - Improbable 4 - Probable 2 - Remote Low Acceptable
SPM electrical components fail
  1. SPM electrical component is of insufficient quality and fails
Operator unable to use SPM Delayed diagnosis 1 - Negligible 1 - Improbable 5 - Frequent 3 - Occasional Low Acceptable
SPM experiences electrostatic discharge
  1. SPM is touched by person carrying electrostatic charge
  2. SPM electronics are damaged by resulting discharge
Operator unable to use SPM Delayed diagnosis 1 - Negligible 1 - Improbable 5 - Frequent 3 - Occasional Low Acceptable
SPM experiences electrical interference
  1. SPM experiences hardware failure due to voltage and electrical noise caused by nearby EM waves
Operator unable to use SPM Delayed diagnosis 1 - Negligible 1 - Improbable 5 - Frequent 3 - Occasional Low Acceptable
SPM is dropped and breakage exposes electrical circuits
  1. SPM is dropped and breakage exposes electrical circuits
Operator exposed to electricity Electrical shock 2 - Minor 2 - Remote 2 - Remote 2 - Remote Low Acceptable
SPM is plugged in to faulty power supply
  1. SPM is plugged in to faulty power supply
  2. SPM electronics overdraw current and catch fire
Operator exposed to fire Burn 3 - Serious 1 - Improbable 2 - Remote 1 - Improbable Low Acceptable
SPM is unplugged during operation on insufficient battery charge
  1. SPM is unplugged during operation on insufficient battery charge
  2. SPM dies during STC operation
  3. STC contains a positive sample that is insufficiently lysed
Patient receives false negative result Incorrect diagnosis 4 - Critical 1 - Improbable 5 - Frequent 3 - Occasional Medium Unacceptable
SPM placed near magnetic field
  1. SPM is placed near magnetic field that triggers lid sensor while lid is open
  2. SPM shakes while exposed
Operator exposed to moving parts Physical trauma 2 - Minor 1 - Improbable 3 - Occasional 2 - Remote Low Acceptable
SPM wires are exposed
  1. SPM internal wiring or power cord insulation breaks down
  2. SPM exposed wiring energizes ungrounded enclosure
  3. SPM enclosure is touched
Operator exposed to electricity Electrical shock 2 - Minor 1 - Improbable 2 - Remote 1 - Improbable Low Acceptable
PM is dropped on person
  1. PM is dropped on person
Person hit by PM Physical trauma 2 - Minor 1 - Improbable 3 - Occasional 2 - Remote Low Acceptable
PM magnetic closure pinches
  1. PM lid is closed on fingers and magnetic closure causes pinch
Operator pinches fingers Physical trauma 2 - Minor 2 - Remote 2 - Remote 2 - Remote Low Acceptable
SPM connector fails during shaking
  1. SPM connecter fails mechanically
  2. SPM is unusable
Operator unable to use SPM Delayed diagnosis 1 - Negligible 1 - Improbable 5 - Frequent 3 - Occasional Low Acceptable
SPM is dropped on person
  1. SPM is dropped on person
Person hit by PM Physical trauma 2 - Minor 1 - Improbable 3 - Occasional 2 - Remote Low Acceptable
SPM loses control of shaker
  1. SPM shaker control fails
  2. SPM shakes even when lid is open
Operator exposed to moving parts Physical trauma 2 - Minor 1 - Improbable 3 - Occasional 2 - Remote Low Acceptable
SPM observation window breaks
  1. SPM observation window breaks
Operator exposed to moving/sharp parts Physical trauma 2 - Minor 1 - Improbable 2 - Remote 1 - Improbable Low Acceptable
SPM magnetic closure pinches
  1. SPM lid is closed on fingers and magnetic closure causes pinch
Operator pinches fingers Physical trauma 2 - Minor 2 - Remote 2 - Remote 2 - Remote Low Acceptable
SPM shakes while open
  1. SPM fails to stop shaking when lid is opened
Operator exposed to moving parts Physical trauma 2 - Minor 1 - Improbable 2 - Remote 1 - Improbable Low Acceptable
SPM vibrates off of surface during shaking
  1. SPM vibrates off surface during shaking and falls on a person
Person hit by PM Physical trauma 2 - Minor 2 - Remote 3 - Occasional 2 - Remote Low Acceptable
System damaged by shipping
  1. System is inadequately packed and is unusable at customer site
Operator unable to use system Delayed diagnosis 1 - Negligible 1 - Improbable 5 - Frequent 3 - Occasional Low Acceptable
STC melts during heating in SPM
  1. STC is operated in SPM
  2. STC melts onto electronics, and they catch fire
Operator exposed to fire Burn 3 - Serious 1 - Improbable 2 - Remote 1 - Improbable Low Acceptable
SPM loses control of heater
  1. SPM heater control fails
  2. SPM heats uncontrollably until it catches fire
Operator exposed to fire Burn 3 - Serious 1 - Improbable 2 - Remote 1 - Improbable Low Acceptable