Risk Evaluation
Overview
The nPOC-BB was originally developed at GHL as the front-end of a prototype diagnostic test system called NAATOS. Therefore, the design inputs for the nPOC-BB are a subset of the design inputs for NAATOS, and the Risk Evaluation for the nPOC-BB is excerpted from the larger NAATOS Risk Evaluation.
You can also download a tabular version of this Risk Evaluation Table in the *.csv format.
Lookup Matrices
Probability
P1 (Probability of the Hazardous Situation Occurring) × P2 (Probability of the Hazardous Situation Leading to Harm) = Probability
| P1 \ P2 | 1 - Improbable | 2 - Remote | 3 - Occasional | 4 - Probable | 5 - Frequent |
|---|---|---|---|---|---|
| 1 - Improbable | 1 - Improbable | 1 - Improbable | 2 - Remote | 2 - Remote | 3 - Occasional |
| 2 - Remote | 1 - Improbable | 2 - Remote | 2 - Remote | 3 - Occasional | 3 - Occasional |
| 3 - Occasional | 2 - Remote | 2 - Remote | 3 - Occasional | 4 - Probable | 4 - Probable |
| 4 - Probable | 2 - Remote | 3 - Occasional | 4 - Probable | 4 - Probable | 5 - Frequent |
| 5 - Frequent | 3 - Occasional | 3 - Occasional | 4 - Probable | 5 - Frequent | 5 - Frequent |
Risk Level
S (Severity) × P (Probability) = Risk Level
| P \ S | 1 - Negligible | 2 - Minor | 3 - Serious | 4 - Critical | 5 - Catastrophic |
|---|---|---|---|---|---|
| 1 - Improbable | Low | Low | Low | Low | Medium |
| 2 - Remote | Low | Low | Medium | Medium | Medium |
| 3 - Occasional | Low | Medium | Medium | Medium | High |
| 4 - Probable | Low | Medium | Medium | High | High |
| 5 - Frequent | Medium | Medium | High | High | High |
Risk Evaluation Table
| Name | Sequence of Events | Hazardous Situation | Harm | Severity | P1 (Probability of the Hazardous Situation Occurring) | P2 (Probability of the Hazardous Situation Leading to Harm) | Ptotal = P1 x P2 | Initial Risk Level | Initial Risk Acceptability |
|---|---|---|---|---|---|---|---|---|---|
| SCC contaminates environment with sample |
|
Person exposed to sample | Disease transmission | 3 - Serious | 2 - Remote | 3 - Occasional | 2 - Remote | Medium | Unacceptable |
| SCC is reused |
|
Person exposed to sample | Disease transmission | 3 - Serious | 1 - Improbable | 5 - Frequent | 3 - Occasional | Medium | Unacceptable |
| STC contaminates environment with sample |
|
Person exposed to sample | Disease transmission | 3 - Serious | 2 - Remote | 3 - Occasional | 2 - Remote | Medium | Unacceptable |
| STC is reused |
|
Person exposed to sample | Disease transmission | 3 - Serious | 1 - Improbable | 1 - Improbable | 1 - Improbable | Low | Acceptable |
| SPM contaminates environment with sample |
|
Person exposed to sample | Disease transmission | 3 - Serious | 2 - Remote | 4 - Probable | 3 - Occasional | Medium | Unacceptable |
| SPM does not inactivate sample |
|
Person exposed to sample | Disease transmission | 3 - Serious | 1 - Improbable | 2 - Remote | 1 - Improbable | Low | Acceptable |
| SCC is contaminated with manufacturing chemicals |
|
Patient exposed to chemicals | Toxicity | 3 - Serious | 1 - Improbable | 2 - Remote | 1 - Improbable | Low | Acceptable |
| SCC is contaminated by environment |
|
Patient receives false positive result | Incorrect diagnosis | 3 - Serious | 1 - Improbable | 4 - Probable | 2 - Remote | Medium | Unacceptable |
| SCC is never added to STC |
|
Patient receives false negative result | Incorrect diagnosis | 4 - Critical | 1 - Improbable | 4 - Probable | 2 - Remote | Medium | Unacceptable |
| STC is contaminated by environment |
|
Patient receives false positive result | Incorrect diagnosis | 3 - Serious | 1 - Improbable | 4 - Probable | 2 - Remote | Medium | Unacceptable |
| STC is opened and left exposed to environment |
|
Operator obtains invalid result | Delayed diagnosis | 1 - Negligible | 1 - Improbable | 3 - Occasional | 2 - Remote | Low | Acceptable |
| STC is removed from SPM during operation |
|
Patient receives false negative result | Incorrect diagnosis | 4 - Critical | 1 - Improbable | 3 - Occasional | 2 - Remote | Medium | Unacceptable |
| STC sample transfer into TC introduces air bubbles |
|
Operator obtains invalid result | Delayed diagnosis | 1 - Negligible | 2 - Remote | 2 - Remote | 2 - Remote | Low | Acceptable |
| STC sample transfer into TC is delayed |
|
Patient receives false negative result | Incorrect diagnosis | 4 - Critical | 1 - Improbable | 2 - Remote | 1 - Improbable | Low | Acceptable |
| STC sample transfer into TC leaks |
|
Operator obtains invalid result | Delayed diagnosis | 1 - Negligible | 1 - Improbable | 3 - Occasional | 2 - Remote | Low | Acceptable |
| STC leaks fluid into SPM |
|
Operator exposed to fire | Burn | 3 - Serious | 1 - Improbable | 2 - Remote | 1 - Improbable | Low | Acceptable |
| SPM batteries overcharge |
|
Operator exposed to fire | Burn | 3 - Serious | 1 - Improbable | 4 - Probable | 2 - Remote | Medium | Unacceptable |
| SPM batteries over-discharge |
|
Operator unable to use SPM | Delayed diagnosis | 1 - Negligible | 1 - Improbable | 4 - Probable | 2 - Remote | Low | Acceptable |
| SPM electrical components fail |
|
Operator unable to use SPM | Delayed diagnosis | 1 - Negligible | 1 - Improbable | 5 - Frequent | 3 - Occasional | Low | Acceptable |
| SPM experiences electrostatic discharge |
|
Operator unable to use SPM | Delayed diagnosis | 1 - Negligible | 1 - Improbable | 5 - Frequent | 3 - Occasional | Low | Acceptable |
| SPM experiences electrical interference |
|
Operator unable to use SPM | Delayed diagnosis | 1 - Negligible | 1 - Improbable | 5 - Frequent | 3 - Occasional | Low | Acceptable |
| SPM is dropped and breakage exposes electrical circuits |
|
Operator exposed to electricity | Electrical shock | 2 - Minor | 2 - Remote | 2 - Remote | 2 - Remote | Low | Acceptable |
| SPM is plugged in to faulty power supply |
|
Operator exposed to fire | Burn | 3 - Serious | 1 - Improbable | 2 - Remote | 1 - Improbable | Low | Acceptable |
| SPM is unplugged during operation on insufficient battery charge |
|
Patient receives false negative result | Incorrect diagnosis | 4 - Critical | 1 - Improbable | 5 - Frequent | 3 - Occasional | Medium | Unacceptable |
| SPM placed near magnetic field |
|
Operator exposed to moving parts | Physical trauma | 2 - Minor | 1 - Improbable | 3 - Occasional | 2 - Remote | Low | Acceptable |
| SPM wires are exposed |
|
Operator exposed to electricity | Electrical shock | 2 - Minor | 1 - Improbable | 2 - Remote | 1 - Improbable | Low | Acceptable |
| PM is dropped on person |
|
Person hit by PM | Physical trauma | 2 - Minor | 1 - Improbable | 3 - Occasional | 2 - Remote | Low | Acceptable |
| PM magnetic closure pinches |
|
Operator pinches fingers | Physical trauma | 2 - Minor | 2 - Remote | 2 - Remote | 2 - Remote | Low | Acceptable |
| SPM connector fails during shaking |
|
Operator unable to use SPM | Delayed diagnosis | 1 - Negligible | 1 - Improbable | 5 - Frequent | 3 - Occasional | Low | Acceptable |
| SPM is dropped on person |
|
Person hit by PM | Physical trauma | 2 - Minor | 1 - Improbable | 3 - Occasional | 2 - Remote | Low | Acceptable |
| SPM loses control of shaker |
|
Operator exposed to moving parts | Physical trauma | 2 - Minor | 1 - Improbable | 3 - Occasional | 2 - Remote | Low | Acceptable |
| SPM observation window breaks |
|
Operator exposed to moving/sharp parts | Physical trauma | 2 - Minor | 1 - Improbable | 2 - Remote | 1 - Improbable | Low | Acceptable |
| SPM magnetic closure pinches |
|
Operator pinches fingers | Physical trauma | 2 - Minor | 2 - Remote | 2 - Remote | 2 - Remote | Low | Acceptable |
| SPM shakes while open |
|
Operator exposed to moving parts | Physical trauma | 2 - Minor | 1 - Improbable | 2 - Remote | 1 - Improbable | Low | Acceptable |
| SPM vibrates off of surface during shaking |
|
Person hit by PM | Physical trauma | 2 - Minor | 2 - Remote | 3 - Occasional | 2 - Remote | Low | Acceptable |
| System damaged by shipping |
|
Operator unable to use system | Delayed diagnosis | 1 - Negligible | 1 - Improbable | 5 - Frequent | 3 - Occasional | Low | Acceptable |
| STC melts during heating in SPM |
|
Operator exposed to fire | Burn | 3 - Serious | 1 - Improbable | 2 - Remote | 1 - Improbable | Low | Acceptable |
| SPM loses control of heater |
|
Operator exposed to fire | Burn | 3 - Serious | 1 - Improbable | 2 - Remote | 1 - Improbable | Low | Acceptable |